Home Product Catalogue Neon & IP Loggers News & Brochures Product Support Where to Buy Connect To Our FTP Contact Us By Phone Contact Us By Email Staff Portal

Product Support

Product Alert #39

8007 MicroLogger

8007 MicroLogger Issues

3 April 1997

39.1 Intermittent SERIN Instruction Operation

Problem

This will only occur when using a Model “A” MicroLogger and the AREF signal as the HSSYNC for HSIO transfers.

The AREF signal (when used as the HSSYNC) on the Model “A” occurs before the HSCLK signal has become stable. This means that an additional clock may be occur, which has the effect of dividing the SERIN reading by two.

Correction

The correction is to delay the AREF signal until after the HSCLK is stable (as is done in the Model “B” units). To do this you must swap the signal controlling the AREF from MPWR to MRUN.

This is done by:-

  1. cutting a track on the top of the PCB
  2. fitting a wire mod under the PCB

Refer to your UNIDATA office for details.

39.2 Scan Synchronised UPS Stays On

Problem

Incorrect operation of the Scan Sync UPS occur in MicroLoggers fitted with Version 60.2 or 61.2 Firmware. Both Models “A” & “B” are affected. (The previous Versions 60.1 and 61.1 did operate correctly.)

Correction

The correction is in Version 60.3 or 61.3 Firmware.