Product Alert #15

STARLOGGER DATA LOGGER - Model 6004A - Serial #s 301 to 750

Starlogger Issues

7 December 1993

15.1 V21.09 Firmware Error – Adjusting Sounder Pitch blanks the Display

Problem

If the ENTER Key is pressed twice in succession whilst adjusting the Sounder Pitch, the Display Contrast is set to 0% and the display goes blank.

Correction

  1. Enter the following key sequence to readjust the Display Contrast:-
    ON, CONFIG, Down key (13 times), Right key (30 times), ENTER
  2. Upgrade the STARLOGGER to V21.10 Firmware.

15.2 Internal Battery monitoring

Problem

Any low-res channel A0 to A7 may now be assigned to monitor the internal battery voltage. In prior revisions, only A6 could be assigned to the battery voltage.

Correction

Use the Keyboard CONFIG command to select the low-res channel, or enter the configuration setting “98 xx” into the Scheme Title.

where xx =
15 – no low-res channel used for battery measurement
16(A0)...23(A7) – selected for battery measurement