STARLOGGER DATA LOGGER - Model 6004A - Serial #s 301 to 750

V21.08 Firmware Error - STARLOGGER Stops with DEAD battery

22 September 1993

Problem

Accidentally connecting the RS-232 Computer cable into the Input Signal connector may cause the STARLOGGER to shutdown.

The large negative voltages (-6V to -12V) that may be present on the computer RS-232 cable, corrupts the internal battery voltage signal so that it reads 0.00V (or a random voltage).

STARLOGGER firmware prior to V21.9 will detect this false reading as a battery fault and immediately enter shutdown mode.

Correction

Upgrade the STARLOGGER to V21.9 Firmware.

Note

  1. NO DAMAGE WILL OCCUR TO THE LOGGER when the cables are incorrectly connected.
  2. Recorded data is protected and may be unloaded normally.
    To restart the logger, do one of the following:
    • Enter the <RESET> command on the keyboard
    • Enter Alt-I in Test Mode (MDL Diagnostics)
    • “Program the Logger with Scheme” (after saving data)
    • Disconnect/reconnect the battery (after saving data).
  3. This problem has been fixed in firmware V21.9 which has been fitted to STARLOGGERs from serial number 751 onwards.