Model: 7420
Technical Specs |
Order Info
These high precision DRUCK transducers use a high integrity silicon diaphragm and are temperature compensated. They connect directly to a Model 7000 MACRO Data Logger (or via a 6142A Universal Amplifier Module to a Model 6004 STARLOGGER).
Each transducer comes factory-calibrated with a calibration certificate. A transducer for a wider calibrated temperature range of -20 to 80°C with an accuracy of ±1.5% of full scale is available on request.

| Accuracy: |
±0.5% (0 to 50°C) |
| Linearity/Hysteresis: |
±0.1% of full scale |
| Operating temperature: |
-20 to 80°C |
| Calibrated temperature: |
0 to 50°C |
| Overpressure: |
3 times rated pressure |
| Pressure connection: |
¼” NPT – flat end |
| Pressure media: |
gases and fluids compatible with silicon and titanium |
| Signal: |
1 analog, 100mV @ 10V excitation |
| Cable: |
4 core, 1 metre |
| Power: |
5V or 10V, 5mA from Logger |
| Size: |
21mm diameter, 75mm long |
| Weight: |
170g |

| Model |
Description |
Range |
| 7420B |
Pressure Transducer,
Gauge Type |
0 to 50kPa |
| 7420C |
0 to 100kPa |
| 7420D |
0 to 200kPa |
| 7420E |
0 to 500kPa |
| 7420F |
0 to 1000kPa |
| 7420G |
0 to 2000kPa |
| 6239 |
Pressure Transducers User Manual Supplement |
|